DH-2000-CAL Deuterium Tungsten Halogen Calibration StandardThe DH-2000-CAL Deuterium Tungsten Halogen Calibration Standard is a UV-NIR light source used to calibrate the absolute spectral response of a radiometric system. With the DH-2000-CAL and SpectraSuite Software, you can determine known absolute intensity values at wavelengths from 220-1050 nm.
The DH-2000-CAL is specifically calibrated for use with optical fibers or a cosine corrector; the calibration data includes absolute intensities for wavelengths between 220-1050 nm at the fiber entrance port for both a bare fiber and a CC-3-UV Cosine Corrector.
Features
-
UV-NIR Calibration Source. For use in performing fast, radiometric calibrations from 220-1050 nm
-
NIST-traceable Calibration from 220-1050 nm. Provides absolute spectral intensity in µW/cm2/nm at the fiber port
-
Calibration Certificate. Calibration data is provided in paper and electronic formats for use with SpectraSuite Software
Included with the Lamp
The DH-2000-CAL comes with the CC-3-UV Cosine Corrector. Also included is a calibration certificate and a diskette with a data file for use with our SpectraSuite Software.
Other Calibration Services
The DH-2000-CAL typically provides about 50 hours of operation before recalibration is necessary. We recalibrate these lamps in-house. (For more information on the DH-2000-RECAL, contact an Applications Scientist.) Also, by using our SPEC-CAL-UV in-house calibration service, it is possible to have your UV spectrometer radiometrically calibrated without purchasing a DH-2000-CAL. The calibration is good for about one year, provided the optical fiber is not removed from the setup, as the system is calibrated specifically for use with a fiber.
User's Tip
Radiometric Calibration Sources are not illumination sources for spectroscopic measurements. For illumination, use the DH-2000.
DH-2000-CAL Spectral Output
These spectral output graphs were taken separately for the deuterium (200-400 nm) and the tungsten halogen bulbs (400-1100 nm) to avoid a sharp line spectrum >400 nm in the deuterium trace.
DH-2000-CAL and Services Ordering Information
Item |
Description |
Price |
DH2000-CAL |
NIST-traceable UV-NIR (220-1050 nm) calibration source |
Call for Price |
DH2000-RECAL |
Recalibration service from 220-1050 nm |
Call for Price |
DH2000-CAL-EXT |
Upgrade for extended range |
Call for Price |
DH2000-RECAL-EXT |
Upgrade recalibration service from 1050-2400 nm |
Call for Price |
DH-2000-CAL Specifications
Dimensions: |
150 mm x 135 mm x 319 mm |
Weight: |
6 kg |
Power consumption: |
25 W (deuterium); 20 W (tungsten halogen) |
Calibrated range: |
220-1050 nm |
Calibration accuracy: |
+/-5% |
Calibration valid for: |
50 hours |
Lamp current: |
Operating 85 V/0.3A |
Lamp voltage: |
350 V |
Power requirements: |
85-264 V 50/60 Hz |
Current voltage drift: |
<0.01% per hour |
Current voltage stability: |
<5 x 10-6 peak-to-peak (0.1-10.0 Hz) |
Humidity: |
5-95% without condensation at 40 °C |
Operating temperature: |
5 °C - 35 °C |
Total power: |
100 W |
Power consumption: |
190 W maximum |
Warm-up time: |
40 minutes (deuterium); 20 minutes (tungsten halogen) |
Markings: |
CE; VDI/VDE 0160; EN 61010 | |