HL-2000-CAL Radiometric Calibration Standard
The HL-2000-CAL Calibrated Tungsten Halogen Light Source can calibrate the absolute spectral response of a radiometric system. With this NIST-traceable light source and SpectraSuite, you can determine known absolute intensity values at wavelengths from 300-1050 nm. The HL-2000-CAL Calibrated Tungsten Halogen Light Source can be used to calibrate the absolute spectral response of a radiometric system. With this NIST-traceable light source and SpectraSuite Software, you can determine known absolute intensity values at wavelengths from 300-1050 nm.
Features
-
NIST-traceable Calibration from 300-1050 nm. Provides absolute spectral intensity in µW/cm2/nm at the fiber port
-
Source can be calibrated specifically for a bare fiber, a fiber plus cosine corrector, or an integrating sphere
-
Calibration Certificate. Calibration data is provided in paper and electronic formats, for use with our our OOIIrrad-C Irradiance Software or SpectraSuite
Calibrated for Use with Fiber and Cosine Corrector
The HL-2000-CAL is designed for calibrating a system consisting of a spectrometer and a cosine corrector. The calibration data for the HL-2000-CAL includes absolute intensities for wavelengths between 300-1050 nm.
Calibrated for Use with Integrating Spheres
The HL-2000-CAL-ISP is designed for calibrating the absolute spectral response of your system when using the ISP-50-8-I Integrating Sphere as your sampling optic. It will also work with the ISP-80-8-1; however, there will be more light loss. The HL-2000-CAL-ISP will not work with the ISP-30-6-1 as the port is too small.
Included with the Lamp
The HL-2000-CAL comes with a regulated 12 VDC power supply. Also included is a calibration certificate and a diskette with data files for use with our irradiance software. A cosine corrector is available for an additional charge .
Pricing
Item Code |
Description |
Price |
HL-2000-CAL |
Radiometric Calibration Standard, 300-1050 nm |
|
HL-2000-CAL-ISP |
Radiometric Calibration Standard, 300-1050 nm |
|
HL-2000-CAL-EXT |
Upgrade for extended range |
|
HL-2000-RECAL |
Radiometric Recalibration Service for the HL-2000-CAL and HL-2000-CAL-ISP |
|
Spectral Output
The typical calibration output when using the HL-2000-CAL with a CC-3 Cosine Corrector.
Specifications
Spectral range (calibrated): |
300-1050 nm (calibrated) |
Power consumption: |
600 mA @ 12 VDC |
Power output: |
6.5 watts |
Bulb life: |
900 hours (recommend recalibration after 50 hours of use) |
Recalibration: |
Required after ~50 hours of operation |
Bulb color temperature: |
3100 K |
Output to bulb: |
5 volts/1.3 amps |
Output regulation: |
0.2% voltage |
Time to stabilized output: |
~30 minutes |
Connector: |
SMA 905 for fiber; 6.35-mm barrel for cosine corrector; PTFE plug for integrating sphere |
|