Radiometric Calibration Standards
|
The LS-1-CAL and HL-2000-CAL Calibrated Tungsten Halogen Light Sources calibrate the absolute spectral response of a radiometric system. With one of these NIST-traceable light sources and our software, you can determine known absolute intensity values at wavelengths from 300-1050 nm.
-
NIST-traceable Calibration from 300-1050 nm. Provides absolute spectral intensity in µW/cm2/nm at the fiber port
-
Sources can be calibrated specifically for a bare fiber, a fiber plus cosine corrector, or an integrating sphere
-
Calibration Certificate. Calibration data is provided in paper and electronic formats. |
|
Calibrated for Use with Fiber and Cosine Corrector
The LS-1-CAL and HL-2000-CAL are specifically designed for calibrating the absolute spectral response of a complete system consisting of a spectrometer and an optical fiber and/or a CC-3-UV Cosine Corrector. The calibration data for both the LS-1-CAL and the HL-2000-CAL includes absolute intensities for wavelengths between 300–1050 nm. |
|
Calibrated for Use with Integrating Sphere
The LS-1-CAL-INT is designed for calibrating the absolute spectral response of your system when using the FOIS-1 Fiber Optic Integrating Sphere as your sampling optic. The LS-1-CAL-INT comes with a PTFE diffuser plug that fits snugly into the sample port of the FOIS-1 Integrating Sphere to measure absolute spectral intensities of LEDs and other emission sources. The HL-2000-CAL-ISP is designed for calibrating the absolute spectral response of your system when using the ISP-50-8-I Integrating Sphere as your sampling optic. |
Included with the Lamps
Each one of these radiometric sources comes with a regulated 12 VDC power supply. Also included is a calibration certificate and a CD with data files for use with our software.
Other Calibration Services
The LS-1-CAL provides 50 hours of operation before recalibration is necessary. We recalibrate these lamps in-house. (For more information on LS-1-RECAL, contact an Applications Scientist.) For those with spectroradiometric systems working below the 300-1050 nm range of the LS-1-CAL and HL-2000-CAL, see our DH-2000-CAL.
Pricing
Item Code |
Description |
Price |
LS-1-CAL |
Radiometric Calibration Standard, 300-1050 nm |
|
LS-1-CAL-INT |
Radiometric Calibration Standard, 300-1050 nm |
|
LS-1-CAL-EXT |
Upgrade for extended range. |
|
LS-1-RECAL |
Radiometric Re-calibration Service for the LS-1-CAL and LS-1-CAL-INT |
|
HL-2000-CAL |
Radiometric Calibration Standard, 300-1050 nm |
|
HL-2000-CAL-ISP |
Radiometric Calibration Standard, 300-1050 nm |
|
HL-2000-RECAL |
Radiometric Re-calibration Service for the HL-2000-CAL and HL-2000-CAL-ISP |
|
LS-1-CAL Spectral Output
Spectral output of the LS-1-CAL using a bare SMA-terminated fiber
Spectral output of the LS-1-CAL using a CC3 cosine corrector
Spectral output of the LS-1-CAL-INT when connected to a FOIS-1 integrating sphere
LS-1-CAL and HL-2000-CAL Specifications
Spectral range (calibrated): |
300-1050 nm (calibrated) |
Power consumption: |
800 mA @ 12 VDC |
Power output: |
6.5 watts |
Bulb life: |
900 hours (recommend recalibration after 50 hours of use) |
Recalibration: |
Required after ~50 hours of operation |
Bulb color temperature: |
3100 K |
Output regulation: |
0.2% voltage |
Time to stabilized output: |
~30 minutes |
Connector: |
SMA 905 for fiber; 6.35-mm barrel for cosine corrector; PTFE plug for integrating sphere | |